Nellcor 595
Monitors Neonates through adults
Convenient selection of alarm limits
User-selectable fast-averaging mode
Onscreen viewing of 48 hours of SpO2, and pulse rate trends
Specifications
Measurement Range
SpO2: 1% to 100%
Pulse Rate: 20 to 250 beats per minute(bpm)
Perfusion Index: 0.03% to 20%
Accuracy
Saturation: (% SpO2:+1 SD)
Without Motion: Adult 70% to 100%+2 Digits, Neonates 70% to 100% +3 digits.
With Motion: Adult and Neonates 70% to 100% +3 digits.
Low Perfusion: 20 to 250 bpm +2 digits.
Dimensions: Weight 5.7 lbs (2.6 Kgs), Size: 3.3” H x 10.4” W x 6.8” D (8.4 cm x 26.4 cm x 17.3 cm).
Displays/Indicators: High Quality LCD graphical screen. Plethysmographic waveform or magnified view.
Visual indicators:pulse search, audible alarms silenced or off, artifact indicator, low battery and battery charging, sensor off, sensor disconnect and SatSeconds clock.
Alarms: Audible and visual alarms for high/low saturation pulse rate.
SatSeconds alarm management settings:10,25,50 and 100, or off.
Audible and visual indicators for low battery.
Audible sensor disconnect alarm.
Displayed saturation and pulse rate alarm limits.
Specifications
Measurement Range
SpO2: 1% to 100%
Pulse Rate: 20 to 250 beats per minute(bpm)
Perfusion Index: 0.03% to 20%
Accuracy
Saturation: (% SpO2:+1 SD)
Without Motion: Adult 70% to 100%+2 Digits, Neonates 70% to 100% +3 digits.
With Motion: Adult and Neonates 70% to 100% +3 digits.
Low Perfusion: 20 to 250 bpm +2 digits.
Dimensions: Weight 5.7 lbs (2.6 Kgs), Size: 3.3” H x 10.4” W x 6.8” D (8.4 cm x 26.4 cm x 17.3 cm).